Bid opportunityClosed
Optical Microscope
COMMERCE, DEPARTMENT OF › NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY › DEPT OF COMMERCE NIST
Response deadlineSep 8, 2026 11:00 AM EDT
View official notice on SAM.gov ↗ (opens in a new tab)
Verify before acting. Users must verify deadlines, amendments, attachments, eligibility, and submission requirements on the official SAM.gov notice.
Read the Data Disclaimer.
Key decision factors
- Response deadline
- Sep 8, 2026 11:00 AM EDT
- Posted
- Sep 2, 2026 12:00 AM EDT
- Notice type
- Solicitation
- Set-aside
- Not provided or not applicable
- PSC
- 6640 — Laboratory Equipment and Supplies
- Place of performance
- Gaithersburg, Maryland
- Current status
- Closed
Notice details
- Official status
- Closed
- Normalized group
- Bid opportunity
- Notice ID
- 9c1ef1e9132e48e39c3fffc46bb1e61f
- Solicitation number
- 1333ND26QNB640568
Description
Displayed as sanitized plain text from SAM.gov. Retrieved Sep 3, 2026 10:24 PM EDT.
Responses to Questions posed by Industry: Q:There is no description of the “imaging modes”. Reflected light Brightfield is assumed, but does the user also want to have Darkfield and DIC? A: Requirements for imaging modes are specified in points c, d, and h of the Technical Specifications, Section III Requirements, page 2 of the Statement of Requirements. Q: A “motorized translational stage” is mentioned in the general description, but not in the requirements. There is no mention of whether the stage is encoded or not, A: The motorized stage must include an integrated measuring system and rotation, as outlined in point i of the Technical Specification, Section III Requirements, page 2 of the Statement of Requirements. Q: It also mentions that the samples will be cleaved from large wafers, but there is no XY size requirement for the stage. A: The size of the XY stage is specified as 100 mm x 100 mm, see point i of the Technical Specification, Section III Requirements, page 2 of the Statement of Requirements. Q: A camera is mentioned without any further specification A: The camera details are specified in point n of the Technical Specifications, Section III Requirements, page 2 of the Statement of Requirements: at least 6 megapixels, a minimum FOV of 20 mm, and a pixel size better than 3.0 µm Q: There is a mention of Image Analysis software but no mention of what the user wishes for it to do. A: The software requirements are specified in the Software section, Section III Requirements, page 2 of the Statement of Requirements. These requirements will satisfy our needs. Q: Encoded capability for the selected imaging modes: A: Requirements for integrated X-Y-Z motorization and encoded capabilities are specified in points g, k, l, and m of the Technical Specifications and in point e of the Software section, Section III Requirements, page 2 of the Statement of Requirements. Q: Issues with the objectives’ specifications. A: We did not specify that our samples are flat. In general, they may come from wafers with less topographical variation, but in some cases, we alter the surface through various processes. Additionally, we may need to create active devices with contacts and wires attached to the cleaved pieces. All of these factors impose constraints on the working distances required for the objectives. These research needs were the main reason for our requirements. Q: If 1mm is sufficient WD for a 100x objective, then why is 1mm not sufficient WD for a 50x objective? A: Exceeding 1mm of WD is not necessary. For our samples, we need a working distance as large as possible (given the numerical aperture) to facilitate easy manipulation and observation. The 50X objective is the one we are most likely to use for our sample inspections. In contrast, we expect to use the 100X objective less frequently, in cases that permit a more restrictive working distance of 1 mm. However, this shorter distance requires extra caution to avoid damaging the samples. SOLICITATION 1333ND26QNB640568 OPTICAL MICROSCOPE This is a solicitation for commercial items prepared in accordance with format in RFO Subpart 12.2, as supplemented with additional information included in this notice. This announcement constitutes the only solicitation; quotations are being requested, and a separate written solicitation document will not be issued. This solicitation is a Request for Quotations (RFQ). The associated North American Industrial Classification System (NAICS) code for this procurement is 334516 – Analytical Laboratory Instrument Manufacturing, with a Small Business size standard of 1,000 employees. This requirement is will be a Full and Open Competition.
Personalized relevance
Would this fit your company?
Create company profile information with a free account to see a deterministic, explainable relevance score. Matching does not determine eligibility or win probability.
Create Free Account
Pursuit risk (separate from your match score)
Document analysis pending
Some solicitation documents have not been analyzed yet. Additional restrictions may still be present.
Pursuit risk is evaluated separately from your match score above and never changes it. This is not a legal or eligibility determination — verify against the official SAM.gov notice and attachments.
Same solicitation, other stages
Related procurement history
Other public MaroNest records that share this exact SAM.gov agency and solicitation number, including presolicitation notices, amendments, and award outcomes. Verify amendment/award relationships on the official SAM.gov notice.